Theory and Practice of Rapid Flow-Through Analysis Based on Optode Detection and Its Application to pH Measurement as a Model Case
- 1 November 1996
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 68 (21) , 3871-3878
- https://doi.org/10.1021/ac951149+
Abstract
Rapid flow-through analysis (RFA) based on optode (optical chemical sensor) detection is proposed, and its performance is discussed using the RFA system equipped with a pH-sensitive optode as a model case. To demonstrate and understand the usefulness of the RFA system, long-lifetime pH-sensitive optodes were prepared using a hydrophilic poly(hydroxyethyl methacrylate) (poly-HEMA) membrane covalently immobilized with a dye containing an amino group such as Congo Red or Nile Blue. The response equation for the optode with the RFA system was proposed, and satisfactory estimation for rapid pH analysis was obtained. The proposed RFA can stand as an advanced method for conventional flow injection analysis.Keywords
This publication has 25 references indexed in Scilit:
- Flow-through type calcium ion selective optodes based on novel neutral ionophores and a lipophilic anionic dyeAnalytica Chimica Acta, 1994
- Optical sensors for a wide pH range based on azo dyes immobilized on a novel supportAnalytica Chimica Acta, 1994
- Indirect Photometric Determination of Potassium Ions with a Bulky Anion and a Crown Ether Using a Flow-Injection Technique.Analytical Sciences, 1994
- Chemically and mechanically resistant carbon dioxide optrode based on a covalently immobilized pH indicatorAnalytica Chimica Acta, 1993
- Lithium ion selective optical fiber sensor based on a novel neutral ionophore and a lipophilic anionic dyeAnalytical Chemistry, 1993
- Comparison of two fibre-optic l-glutamate biosensors based on the detection of oxygen or carbon dioxide, and their application in combination with flow-injection analysis to the determination of glutamateAnalytica Chimica Acta, 1991
- Analytical performance of an optical pH sensor for acid-base titrationAnalytica Chimica Acta, 1990
- Flow injection of lithium ion using chromogenic 14-crown-4 derivatives as extraction-spectrophotometric reagentsThe Analyst, 1990
- Flow injection analysis as a diagnostic technique for development and testing of chemical sensorsAnalytica Chimica Acta, 1988
- Flow injection analysis : Part IX. A New Approach to Continuous Flow TitrationsAnalytica Chimica Acta, 1977