Testing the safety of baxter continuous cardiac output monitoring system
- 1 May 1996
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 12 (3) , 243-249
- https://doi.org/10.1007/bf00857646
Abstract
The safety of a new continuous cardiac output monitoring system, recently introduced by Baxter Healthcare Corporation's Edwards Critical-Care Division, was evaluated in normal sheep. The study compared the biocompatibility and safety of the Vigilance® CCO Monitoring System, which employs a continuous cardiac output (CCO) catheter with Baxter Edwards' standard Paceport™ pulmonary artery catheter. The CCO catheter, which monitors hemodynamic pressures and provides continuons measurement of cardiac output based on the thermodilution principle, contains a thermal filament that is powered and controlled by a unique cardiac output monitor. Parameters were measured periodically in conscious animals and complete necropsies were performed alter each study. Time Control, Paceport™, and four CCO groups were studied. Selected groups were studied for 3 days (acute), 7 days (subacute), and/or 4 weeks after 3 days of continuons use (recovery). Results showed no significant differences between the CCO and Paceport™ catheters in any of the parameters studied. On gross pathology, observations were similar. The only difference between catheters were microscopie findings of focal subendothelial or subendocardial changes correlated with areas that could have corne into contact with the CCO catheter. In acute groups, these changes consisted of a localized myofiber degeneration or necrosis, while in subacute and recovery groups, consisted only of fibrosis. None of the changes were clinically significant. Thus, the CCO catheter, used in conjunction with the Vigilance® CCO Monitoring System, appears to pose no additional risk over a standard Paceport™ catheter in normal sheep after continuons use for up to 7 days.Keywords
This publication has 8 references indexed in Scilit:
- A Safety Comparison Between A New Continuous Cardiac Output (CCO) Monitoring System And A Standard Pulmonary Artery Catheter In SheepAnesthesiology, 1992
- EVALUATION OF CONTINUOUS CARDIAC OUTPUT IN PATIENTS UNDERGOING CORONARY ARTERY BYPASS SURGERYAnesthesiology, 1992
- Continuous thermodilution cardiac output measurement in intensive care unit patientsJournal of Cardiothoracic and Vascular Anesthesia, 1992
- Comparison of catheter ablation using radiofrequency versus direct current energy: Biophysical, electrophysioiogic and pathologic observationsJournal of the American College of Cardiology, 1991
- A continuous cardiac output computer based on thermodilution principlesAnnals of Biomedical Engineering, 1989
- Continuous Thermal Measurement of Cardiac OutputIEEE Transactions on Biomedical Engineering, 1984
- Cardiac output estimation by a thermodilution method involving intravascular heating and thermistor recording.1970
- DETERMINATION OF CARDIAC OUTPUT IN MAN BY A NEW METHOD BASED ON THERMODILUTIONThe Lancet, 1963