Warping of cross sections in the torsion of long bones with internal fracture fixation plates
- 1 January 1987
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 5 (2) , 296-299
- https://doi.org/10.1002/jor.1100050216
Abstract
Torsion of noncircular beams results in warping of each cross section. When noncircular cross sections are constrained to remain plane, the resulting shear stress distribution is different from what Saint Venant torsion (with warping) would predict. This has practical implications to the stress analysis of plated long bones subjected to torsional loadings. Analyses in which warping is not allowed predict incorrect stress fields in the plate and bone and overpredict the amount of stress shielding associated with fracture plate fixation.Keywords
This publication has 0 references indexed in Scilit: