European Concerted Action on Anticoagulation (ECAA): International Normalized Ratio Variability of CoaguChek and TAS Point-of-Care Testing Whole Blood Prothrombin Time Monitors
- 1 January 2002
- journal article
- schattauer gmbh
- Published by Georg Thieme Verlag KG in Thrombosis and Haemostasis
- Vol. 88 (12) , 992-995
- https://doi.org/10.1055/s-0037-1613345
Abstract
The object was to assess the variability in displayed International Normalised Ratio (INR) between monitors of the same manufacture using whole blood samples from the same subjects. Two brands of monitor, CoaguChek Mini and the TAS PT-NC were tested. 14 instruments of each brand were tested on the same day at the same laboratory by the same operator using identical blood samples to avoid between-centre differences in samples and operator technique. Whole blood samples from two normal donors and four coumarintreated patients were tested to assess between-instrument variability of INR. Results have been coded. There was a much wider dispersion of INR on Brand B than on Brand A. One Brand A instrument failed to give a result with one of the two whole blood samples from one patient. One Brand B monitor gave an aberrant result with one of the samples from a normal subject. On both brands of monitor, INR variability appeared to be due mainly to duplication differences rather than between-instrument variability on both normal and coumarin whole blood samples. * ECAA Steering Group MembersKeywords
This publication has 6 references indexed in Scilit:
- Monitoring oral anticoagulant treatment with the TAS near-patient test system: comparison with conventional thromboplastins.Journal of Clinical Pathology, 1997
- Advances in the monitoring of oral anticoagulation: Point-of-care testing, patient self-monitoring, and patient self-managementJournal of Thrombosis and Thrombolysis, 1996
- The Value of Plasma Calibrants in Correcting Coagulometer Effects on International Normalized Ratios:An International Multicenter StudyAmerican Journal of Clinical Pathology, 1995
- Home Prothrombin Time Monitoring after the Initiation of Warfarin TherapyAnnals of Internal Medicine, 1989
- Comparison of Two Automated Coagulometers and the Manual Tilt-Tube Method for the Determination of Prothrombin TimeAmerican Journal of Clinical Pathology, 1989
- Effect of automation on prothrombin time test in NEQAS surveys.Journal of Clinical Pathology, 1989