Beating Heart Catheter-Based Edge-to-Edge Mitral Valve Procedure in a Porcine Model

Abstract
Background— Surgical edge-to-edge repair has been used in the treatment of mitral regurgitation. We evaluated the ability of a catheter-delivered clip (Evalve, Inc) to achieve edge-to-edge mitral valve approximation without cardiopulmonary bypass and the healing response of this technique. Methods and Results— Twenty-one pigs underwent general anesthesia and left thoracotomy. A 10F flexible delivery catheter with a clip was placed into the left atrium. With echocardiographic and fluoroscopic guidance, the clip grasped and approximated the mid portion of the anterior and posterior leaflets. After a double orifice had been confirmed, the clip was detached and the catheter withdrawn. All animals survived and had successful clip placement. Three animals were euthanized at 4 weeks, 9 at 12 weeks, 1 at 17 weeks, 7 at 24 weeks, and 1 at 52 weeks. The clip was well positioned, with leaflet approximation in all animals except 1, in which the clip separated from the posterior leaflet at 4 weeks without affecting va...