Compliance effects on small diameter polyurethane graft patency
- 1 October 1993
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 27 (10) , 1269-1279
- https://doi.org/10.1002/jbm.820271007
Abstract
Microporous compliance matched and noncompliant grafts were compared in a dog carotid artery interposition model. We fabricated 4 mm diameter sponge type polyurethane (Biomer) tubes 5 cm in length with a 0.5 mm wall thickness. The luminal surface was covered with a 50 μm coating of cross‐linked gelatin. Compliance was measured in vitro and in vivo by volume and vessel diameter changes. Over a mean arterial pressure range of 55–155 mm Hg, the diameter changes of grafts and stump arteries were measured in situ using an ultrasonic Hokanson device. Compliance matched grafts were found to have the same in vitro compliance values as the natural canine carotid at a mean arterial pressure of 100 mm Hg. Compliance matched and noncompliant grafts had values of 10.3 ± 1.3 and 0.9 ± 0.1 × 10−2 mm Hg, respectively. End to end arterial anastomoses were constructed between the graft and the host arteries.The use of synthetic grafts with matched compliance to the adjacent natural vessels has been advocated as the ideal solution to circumvent the problems of graft failure. These studies indicate that compliance values for compliance matched grafts decreased immediately after implantation (from 10.3 to 6.5 × 10−2 %/mm Hg) and within 6 weeks decreased to 3.6 × 10−2 %/mm Hg. The compliance values for noncompliant grafts remained constant throughout the test period.At autopsy all grafts showed a tightly adhered tissue capsule. The thickness of the anastomotic hyperplasia at the distal sites of compliance matched grafts was significantly different (P < .05) than that of the adjacent artery. The patency for compliant and noncompliant grafts was 64% and 50%, respectively. Evidence for polyurethane graft degradation was obtained by Fourier transform infrared spectroscopy and gel permeation chromatography analysis of patent explants. Compliance mismatch alone does not contribute to graft failure, however, material degradation, suture technique and/or capsule formation can play a contributory role although these were not tested directly. © 1993 John Wiley & Sons, Inc.Keywords
This publication has 23 references indexed in Scilit:
- Small‐caliber polyurethane and polytetrafluoroethylene grafts: A comparative study in a canine aortoiliac modelJournal of Biomedical Materials Research, 1990
- Graft Compliance and Anastomotic Flow PatternsAnnual Northeast Bioengineering Conference, 1990
- Establishment of a neovascular bed in a collegen‐impregnated polyurethane spongeJournal of Biomedical Materials Research, 1989
- Analysis of in vitro enzymatic and oxidative degradation of polyurethanesJournal of Biomedical Materials Research, 1988
- Long term biodegradation in vitro of poly(ether urethane urea): a mechanical property studyPolymer, 1987
- Viscoelastic behavior of polyurethane vascular prosthesesJournal of Biomedical Materials Research, 1987
- Physicochemical characterization of a hydrophilic microporous polyurethane vascular graftJournal of Biomedical Materials Research, 1987
- Autooxidative Degradation of Implanted Polyether Polyurethane DevicesJournal of Biomaterials Applications, 1986
- A novel process for the manufacturing of porous grafts: Process description and product evaluationJournal of Biomedical Materials Research, 1983
- Transcutaneous Measurement of the Elastic Properties of the Human Femoral ArteryCirculation, 1972