On the automated definition of mobile target volumes from 4D‐CT images for stereotactic body radiotherapy
- 26 October 2005
- journal article
- radiation therapy-physics
- Published by Wiley in Medical Physics
- Vol. 32 (11) , 3493-3502
- https://doi.org/10.1118/1.2106448
Abstract
Stereotactic body radiotherapy (SBRT) can be used to treat small lesions in the chest. A vacuum‐based immobilization system is used in our clinic for SBRT, and a motion envelope is used in treatment planning. The purpose of this study is to automatically derive motion envelopes using deformable image registration of 4D‐CT images, and to assess the effect of abdominal pressure on the motion envelopes. 4D‐CT scans at ten phases were acquired prior to treatment for both free and restricted breathing using a vacuum‐based immobilization system that includes an abdominal pressure pillow. To study the stability of the motion envelope over the course of treatment, a mid‐treatment 4D‐CT scan was obtained after delivery of the third fraction for two patients. The planning target volume excluding breathing motion was defined on the image set at full exhalation phase and transformed into all other phases using displacement maps from deformable image registration. The motion envelope was obtained as the union of masks of all phases. The ratios of the motion envelope to volume ranged from 1.3 to 2.5. When pressure was applied, the ratios were reduced by as much as 29% compared to free breathing for some patients, but increased by up to 9% for others. The abdominal pressure pillow has more motion restriction effects on the anterior/inferior region of the lung. For one of the two patients for whom the 4D‐CT scan was repeated at mid‐treatment, the motion envelope was reproducible. However, for the other patient the tumor location and lung motion pattern significantly changed due to changes in the anatomy surrounding the tumor during the course of treatment, indicating that an image‐guided approach to SBRT may increase the efficacy of this treatment.Keywords
This publication has 23 references indexed in Scilit:
- A challenge to traditional radiation oncologyInternational Journal of Radiation Oncology*Biology*Physics, 2004
- Treatment plan optimization incorporating respiratory motionMedical Physics, 2004
- 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CTMedical Physics, 2004
- A method for the reconstruction of four‐dimensional synchronized CT scans acquired during free breathingMedical Physics, 2003
- Breathing‐synchronized radiotherapy program at the University of California Davis Cancer CenterMedical Physics, 2000
- Estimation of tumor control probability model parameters from 3-D dose distributions of non-small cell lung cancer patientsLung Cancer, 1999
- Lung cancer: intermittent irradiation synchronized with respiratory motion--results of a pilot study.Radiology, 1998
- Improved Survival in Stage III Non-Small-Cell Lung Cancer: Seven-Year Follow-up of Cancer and Leukemia Group B (CALGB) 8433 TrialJNCI Journal of the National Cancer Institute, 1996
- Randomised multicentre trials of CHART vs conventional radiotherapy in head and neck and non-small-cell lung cancer: an interim report. CHART Steering CommitteeBritish Journal of Cancer, 1996
- Radiotherapy Alone Versus Combined Chemotherapy and Radiotherapy in Nonresectable Non-Small-Cell Lung Cancer: First Analysis of a Randomized Trial in 353 PatientsJNCI Journal of the National Cancer Institute, 1991