Potential problems with the random-zero sphygmomanometer.
- 1 February 1994
- journal article
- abstracts
- Published by Wolters Kluwer Health in Hypertension
- Vol. 23 (2) , 254-257
- https://doi.org/10.1161/01.hyp.23.2.254
Abstract
The random-zero sphygmomanometer is frequently used in epidemiologic and clinical research to increase the reliability and validity of blood pressure readings. However, recent reports have suggested that there is a correlation between the zero values of a random-zero sphygmomanometer and the zero-corrected blood pressure readings obtained. The design of the random-zero sphygmomanometer would suggest that the zero values and zero-corrected blood pressures should be uncorrelated. Hence, residual correlation might be of importance in determining the utility of this device. We have explored this relation in the Middlesex County Blood Pressure Survey, which collected data on 2725 randomly selected adults. Each person had three blood pressure readings taken with a random-zero sphygmomanometer operated by trained interviewers. There was a very weak but statistically significant correlation between the zero values and zero-corrected systolic blood pressure (r = .034, regression slope = -.10), but there was no statistically significant relation with zero-corrected diastolic blood pressure (r = .0003, slope = .0006). Both the correlations and regression slopes were higher for subjects over age 65 years. These data fail to confirm the observed correlations found by Kronmal et al. This discrepancy might be explained by differences in measurement technique, which could introduce a blood pressure-dependent skewing of the range of zero values. If confirmed, this effect would have no effect on the validity of the final blood pressure readings and hence would not need to be considered in decisions about the use of the random-zero sphygmomanometer.Keywords
This publication has 13 references indexed in Scilit:
- Properties of the random zero sphygmomanometer.Hypertension, 1993
- Ambulatory blood pressure monitoring use in hypertension research and clinical practice.Hypertension, 1993
- The cardiovascular health study: Design and rationaleAnnals of Epidemiology, 1991
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- Potential problems with random zero sphygmomanometerThe Lancet, 1990
- A comparison of the random-zero and standard mercury sphygmomanometers.Hypertension, 1988
- Training and certification of blood pressure observers.Hypertension, 1983
- Evaluation of performance of selected devices for measuring blood pressureThe American Journal of Cardiology, 1973
- A RANDOM-ZERO SPHYGMOMANOMETERThe Lancet, 1970
- THE NATURE OF ESSENTIAL HYPERTENSIONThe Lancet, 1960