Current thresholds and liminal size in excitation of heart muscle
- 1 August 1978
- journal article
- research article
- Published by Oxford University Press (OUP) in Cardiovascular Research
- Vol. 12 (8) , 477-485
- https://doi.org/10.1093/cvr/12.8.477
Abstract
End-diastolic current thresholds have been measured in 12 open-chested dogs with epicardial disc electrodes ranging in radius from 0.01 to 9.0 mm. The current threshold for cathodal 1 ms impulses proved to be independent of electrode size for radii smaller than 0.2 mm and proportional to the radius to the power 1.5 for radii greater than 0.4 mm. This relationship can be explained on the assumption that excitation occurs if the current passing through a ‘liminal area’ with dimensions of about 0.3 mm exceeds a critical value. This result is in agreement with the liminal length concept used to describe excitation of cardiac Purkinje fibres. The existance of a liminal area or liminal length for excitation implies that there will be a limit to increasing stimulation efficiency by applying smaller stimulation electrodes.Keywords
This publication has 4 references indexed in Scilit:
- Directional differences of impulse spread in trabecular muscle from mammalian heart.The Journal of Physiology, 1976
- Myocardial stimulation: the electrochemistry of electrode-tissue couplingAmerican Journal of Physiology-Legacy Content, 1967
- Resistivity of Body Tissues at Low FrequenciesCirculation Research, 1963