Venous assessment using air plethysmography: A comparison with clinical examination, ambulatory venous pressure measurement and duplex scanning
- 1 August 1993
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in British Journal of Surgery
- Vol. 80 (8) , 967-970
- https://doi.org/10.1002/bjs.1800800808
Abstract
Air plethysmography was compared with clinical assessment, ambulatory venous pressure measurement and duplex ultrasonography in 103 unselected limbs with venous disease and ten normal control limbs without such disease. Measurements of venous function obtained by air plethysmography showed considerable overlap between groups of limbs classified on the basis of clinical condition or by the presence of popliteal incompetence detected by duplex scanning. The measurement of venous refilling time using air plethysmography correlated poorly with that obtained by venous cannulation (rs = 0·58). The residual volume fraction did not correlate with ambulatory venous pressure measurement (rs = 0·04). Air plethysmography was not found to be as useful as previously reported. The residual volume fraction should not be accepted as a substitute for ambulatory venous pressure measurement, which remains the ‘gold standard’ test of venous function.Keywords
This publication has 3 references indexed in Scilit:
- Quantitative segmental evaluation of venous valvular reflux with duplex ultrasound scanningJournal of Vascular Surgery, 1989
- Venous reflux: Quantification and correlation with the clinical severity of chronic venous diseaseBritish Journal of Surgery, 1988