A Three-Axis Hip Joint Simulator for Wear and Friction Studies on Total Hip Prostheses
- 1 September 1996
- journal article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 210 (3) , 175-185
- https://doi.org/10.1243/pime_proc_1996_210_410_02
Abstract
A three-axial, single-station hip joint simulator was designed and built for wear and friction studies on total hip prostheses. The design of the apparatus is described in detail. Continuous level walking is simulated. All three motion components, flexion-extension, abduction-adduction and internal-external rotation, are included. The motions are implemented electromechanically and the uniaxial load pneumatically. The load is measured continuously. For accurate measurement of wear, the apparatus has a loaded control joint, which also renders both the test and control joints self-centring, as they are loaded in series. The frictional torque of the test joint can be measured continuously throughout the wear test, which is an exceptional feature. Four tests of five million cycles each were completed using 32 mm diameter Co-Cr-Mo femoral heads and 5.6 mm thick, metal-backed, ultra-high molecular weight polyethylene acetabular cups as test specimens. Their wear and friction behaviour is described and discussed in relation to previous simulator studies and clinical observations. The lubricant was distilled water, maintained at body temperature. The wear of the cups was measured gravimetrically at intervals. The average wear rate was 3.9 mg/one million cycles, corresponding to 0.03 mm/year, and the average coefficient of friction was 0.01.Keywords
This publication has 25 references indexed in Scilit:
- An evaluation of the penetration of ceramic femoral heads into polyethylene acetabular cupsWear, 1993
- Wear of polyethylene acetabular cups against alumina femoral heads: 5 prostheses compared in a hip simulator for 35 million walking cyclesActa Orthopaedica, 1993
- Wear of the polyethylene acetabular cup: Metallic and ceramic heads compared in a hip simulatorActa Orthopaedica, 1993
- A Simulator Study of Friction in Total Replacement Hip JointsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1992
- A Five-Station Hip Joint Simulator for Wear Rate StudiesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1992
- Soft layered prostheses for arthritic hip joints: a study of materials degradationJournal of Biomedical Engineering, 1991
- Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: III: The measurement of internal volume changes in explanted Charnley sockets after 2–16 years in vivo and the determination of wear factorsWear, 1985
- Comparative laboratory studies on the production of soluble and particulate metal by total joint prosthesesArchives of orthopaedic and trauma surgery, 1979
- A comparative study of total hip replacement prosthesesJournal of Biomechanics, 1973
- The tribology (friction, lubrication and wear) of all-metal artificial hip jointsWear, 1971