Titration of Power Output During Radiofrequency Catheter Ablation of Atrioventricular Nodal Reentrant Tachycardia

Abstract
Radiofrequency lesions in the anterior, superior aspect of the tricuspid annulus result in selective elimination of fast pathway function in patients with typical atrioventricular (AV) nodal reentry tachycardia. This technique is simple and effective, but has been associated with a significant risk of inadvertent complete AV block. The purpose of this study was to compare the safety and effectiveness of two different techniques for radiofrequency catheter ablation of the fast AV nodal pathway. Initially, a fixed power output was used at each target site. This method was compared retrospectively to a newer technique where power output was gradually incremented at each site. Radiofrequency power was initially applied at 10 watts for 10-15 seconds. If no junctional ectopy or a change in PR interval was seen, power output was incremented by 2 to 4 watts every 10 to 15 seconds up to a maximum of 30 watts. Thirty seven of 38 (96%) patients treated using this incremental power output were cured of their AV nodal reentry tachycardia. None of these patients developed inadvertent complete AV block. In contrast, 92% of historic controls treated with a fixed power output between 20 and 30 watts achieved a primary success and nine of these 89 (10%) historic controls developed inadvertent complete AV block (P = 0.04). There was no difference in the amplitudes of atrial, His, or ventricular electrograms at the effective sites between the two groups.(ABSTRACT TRUNCATED AT 250 WORDS)

This publication has 9 references indexed in Scilit: