Computer controlled ENDOR spectrometer
- 1 July 1974
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 7 (7) , 569-574
- https://doi.org/10.1088/0022-3735/7/7/021
Abstract
The design and construction of a computer-controlled X-band ENDOR spectrometer are described. The instrument operates in the RF band 3-39 MHz with a maximum power of 1 kW, corresponding to fields of 80 G peak to peak in the cavity. Details of a new construction for the microwave cavity are given. A minicomputer is used for control of the experiment and data acquisition, processing and display. Examples of the performance are presented.Keywords
This publication has 5 references indexed in Scilit:
- Versatile interface for PDP-8 computersJournal of Physics E: Scientific Instruments, 1973
- Electron Nuclear Double ResonanceAnnual Review of Physical Chemistry, 1971
- ENDOR of Free Radicals in SolutionThe Journal of Chemical Physics, 1965
- Variable Coupling Reflection Cavity for Microwave SpectroscopyReview of Scientific Instruments, 1961
- Polarization of Phosphorus Nuclei in SiliconPhysical Review B, 1956