The retention mechanics of LTI carbon, carbon‐coated aluminum oxide, and uncoated aluminum oxide dental implants
- 13 September 1983
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 17 (5) , 873-883
- https://doi.org/10.1002/jbm.820170514
Abstract
The in vitro mechanical behavior of identical geometry LTI carbon, carbon‐coated aluminum oxide, and uncoated aluminum oxide blade‐type dental implants has been evaluated using rosette type strain gauges and a LVDT system. The implants served as a distal abutment for a three‐unit fixed prosthesis and functioned for a period of 2 years in female baboons. The comparison of the LTI carbon implants to the carbon‐coated aluminum oxide implants allowed for a study of the effect of implant elastic modulus on the mechanical behavior, while the comparison of the carbon‐coated and uncoated aluminum oxide implants allowed for a study of the effect of chemical composition at the tissue–implant interface. The results of the mechanical testing indicate that the implant displacement response of the LTI carbon implants was greater than that of the carbon‐coated and uncoated aluminum oxide implants. Little difference in displacement response was observed for the carbon‐coated and uncoated aluminum oxide implants. No clear trend in the strain response of the buccal mandibular bone was observed for the three implant systems. The greatest differences in strain response, however, was observed between the uncoated aluminum oxide implants and the LTI carbon and carbon‐coated aluminum oxide implants.This publication has 12 references indexed in Scilit:
- Quantitative histologic evaluation of LTI carbon, carbon‐coated aluminum oxide and uncoated aluminum oxide dental implantsJournal of Biomedical Materials Research, 1983
- A clinical comparison of LTI carbon, alumina, and carbon‐coated alumina blade‐type implants in baboonsJournal of Biomedical Materials Research, 1982
- Retention Characteristics of Porous Rooted Co-Cr-Mo Alloy Dental ImplantsBiomaterials, Medical Devices, and Artificial Organs, 1982
- The influence of implant elastic modulus on the stress distribution around LTI carbon and aluminum oxide dental implantsJournal of Biomedical Materials Research, 1981
- An evaluation of ion‐textured aluminum oxide dental implantsJournal of Biomedical Materials Research, 1981
- Strain gage application on compact boneJournal of Biomechanics, 1979
- Stress Analysis of Porous Rooted Dental ImplantsJournal of Dental Research, 1976
- Interaction effects among cortical bone, cancellous bone, and periodontal membrane of natural teeth and implantsJournal of Biomedical Materials Research, 1976
- Load-Bearing Capacity of Functioning Alumina Dental Endosseous ImplantsJournal of Dental Research, 1976
- Biomechanical and Biomaterial Considerations of Natural Teeth, Tooth Replacements, and Skeletal FixationBiomaterials, Medical Devices, and Artificial Organs, 1974