Modular Tibial Insert Micromotion
- 1 November 1998
- journal article
- symposium
- Published by Wolters Kluwer Health
- Vol. 356, 10-15
- https://doi.org/10.1097/00003086-199811000-00003
Abstract
This study investigates micromotion between modular tibial components, one of the causes of wear on the undersurface of polyethylene inserts. The authors measured motion at the interface of nine contemporary total knee implant designs by mechanically testing the implants in a servohydraulic testing machine. Anteroposterior and mediolateral motion between the tibial insert and baseplate were measured with an extensometer placed across the interface. These tests revealed that in all implants analyzed, sufficient motion occurred to create fretting at the modular interface. Although the testing configuration in this study was not a simulation of in situ loading patterns, the authors observed hundreds of microns of motion even under a 100 N load and variability between implants of the same design, showing that there is room for improvement among locking mechanisms in modular total knee implants.Keywords
This publication has 5 references indexed in Scilit:
- Polyethylene wear on the distal tibial insert surface in total knee arthroplastyThe Knee, 1998
- Tibial Insert Undersurface as a Contributing Source of Polyethylene Wear DebrisClinical Orthopaedics and Related Research, 1997
- Failure of the porous-coated anatomic prosthesis in total knee arthroplasty due to severe polyethylene wear.Journal of Bone and Joint Surgery, 1993
- Bone lysis in well-fixed cemented femoral componentsThe Journal of Bone and Joint Surgery. British volume, 1990
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977