Double inversion black‐blood fast spin‐echo imaging of the human heart: A comparison between 1.5T and 3.0T
Open Access
- 19 May 2003
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 17 (6) , 648-655
- https://doi.org/10.1002/jmri.10316
Abstract
Purpose: To evaluate the effectiveness of blood suppression and the quality of black‐blood cardiac images acquired at 3.0 Tesla using a double‐inversion recovery fast spin‐echo sequence by comparing data acquired at 3.0T to data acquired at 1.5T.Materials and Methods: Black‐blood T2‐weighted fast spin‐echo images of the heart were acquired from five normal volunteers at 1.5T and five normal volunteers at 3.0T. Region‐of‐interest signal intensity measurements were performed at several locations in the suppressed blood regions of the left and right ventricles and around the left ventricle walls to assess the effectiveness and uniformity of the blood suppression, the myocardial signal‐to‐noise ratio (SNR), and the signal uniformity at both field strengths. B1 field maps were produced in phantoms and in subjects at both field strengths.Results: Blood suppression performance is equivalent at 1.5T and 3.0T. The improvement in SNR at 3.0T compared with 1.5T is less than has been predicted in previous studies. The signal uniformity is significantly poorer at 3.0T than at 1.5T due to dielectric effects and shorter radio frequency wavelengths (P < 0.005).Conclusion: Spin‐echo and spin‐echo echo‐train sequences that perform well at 1.5T will produce large signal variations in the chest cavity at 3.0T without modifications. B1 insensitive methods must be explored and implemented for spin‐echo sequences to fully realize the advantages of using these sequences for high‐field MRI. J. Magn. Reson. Imaging 2003;17:648–655.Keywords
This publication has 20 references indexed in Scilit:
- The Intrinsic Signal-to-Noise Ratio in Human Cardiac Imaging at 1.5, 3, and 4 TJournal of Magnetic Resonance, 1997
- Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)IEEE Transactions on Medical Imaging, 1991
- A phase-cycling technique for canceling spurious echoes in NMR imagingJournal of Magnetic Resonance (1969), 1987
- Comparison of linear and circular polarization for magnetic resonance imagingJournal of Magnetic Resonance (1969), 1985
- Estimating radiofrequency power deposition in body NMR imagingMagnetic Resonance in Medicine, 1985
- An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 TJournal of Magnetic Resonance (1969), 1985
- MR ImagingJournal of Computer Assisted Tomography, 1985
- RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imagingPhysics in Medicine & Biology, 1978
- Suppression of artifacts in two-dimensional J spectroscopyJournal of Magnetic Resonance (1969), 1977
- The signal-to-noise ratio of the nuclear magnetic resonance experimentJournal of Magnetic Resonance (1969), 1976