The Elevated-Rim Acetabular Liner in Total Hip Arthroplasty
- 1 January 1996
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 78 (1) , 80-6
- https://doi.org/10.2106/00004623-199601000-00011
Abstract
Although an acetabular component with an elevated rim is thought to improve the postoperative stability of a total hip prosthesis, the actual clinical value has not yet been demonstrated. To address this question, we reviewed the results of 5167 total hip arthroplasties that had been performed at our institution from April 1, 1985, through December 31, 1991. The prostheses included 2469 acetabular components with an elevated-rim liner (10 degrees of elevation) and 2698 with a standard liner. The cumulative probability of dislocation was estimated as a function of time since the operation with use of the Kaplan-Meier survivorship method. Forty-eight of the 2469 hips that had the elevated-rim acetabular liner dislocated within two years, compared with 101 of the 2698 hips that had the standard acetabular liner. The two-year probability of dislocation was 2.19 per cent for the hips with the elevated-rim liner and 3.85 per cent for those with the standard liner (p = 0.001). A similar trend was seen at five years; however, because of a smaller sample the difference was not significant. Increased stability at two years was also demonstrated for the hips with the elevated-rim liner when the hips were analyzed according to the operative approach, the mode of fixation, the sex of the patient, and the type of total hip arthroplasty (primary or revision). Although these data demonstrate improved stability after total hip arthroplasty when an elevated liner is used, particularly in hips that are at greater risk for dislocation of the prosthesis, the long-term effect of this elevated liner on wear and loosening remains unknown but is of considerable concern. The elevated liner deserves additional study to clarify its effect on wear and loosening.Keywords
This publication has 10 references indexed in Scilit:
- Loosening of a femoral stem associated with the use of an extended-lip acetabular cup liner: A case reportThe Journal of Arthroplasty, 1993
- Impingement and loosening of the long posterior wall acetabular implantThe Journal of Bone and Joint Surgery. British volume, 1992
- Dislocation of total hip replacements. A comparative study of standard, long posterior wall and augmented acetabular componentsThe Journal of Bone and Joint Surgery. British volume, 1990
- Failure of acetabular augmentation for recurrent dislocation after hip arthroplasty Report of 3 casesActa Orthopaedica, 1989
- Recurrent dislocation following hip replacement: brief reportThe Journal of Bone and Joint Surgery. British volume, 1988
- Socket wall addition for dislocating total hip: Report of two casesActa Orthopaedica, 1986
- Recurrent dislocation after total hip replacement. Treatment by fixing an additional sector to the acetabular componentThe Journal of Bone and Joint Surgery. British volume, 1985
- Dislocations after total hip arthroplasty.Journal of Bone and Joint Surgery, 1982
- Asymptotically Efficient Rank Invariant Test ProceduresJournal of the Royal Statistical Society. Series A (General), 1972
- Nonparametric Estimation from Incomplete ObservationsJournal of the American Statistical Association, 1958