A feasibility study for megavoltage cone beam CT using a commercial EPID
- 1 January 1998
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 43 (1) , 155-169
- https://doi.org/10.1088/0031-9155/43/1/010
Abstract
This study used a standard commercial electronic portal imaging device (EPID) area detector attached to an isocentric linear accelerator and the Feldkamp algorithm to produce cone beam tomographic reconstructions. The EPID has a active area of 32.5 x 32.5 cm2, and can record 12-bit images using two monitor units (MU), with a resolution of 2.1 x 2.0 mm2 FWHM. Since the EPID was not large enough to record the full patient projection at about 1.5 geometric magnification, it was necessary to offset the detector to collect half-cone projections. Corrections are required to convert pixel values into units of exit dose and to realign the projections to overcome the +/- 4 mm support arm sag. With a geometric magnification of 1.5 the sensitive volume is a cylinder of radius 21 cm and length 17 cm. Unfortunately, the patient couch contains metal bed support rails that lie just outside this cylinder, and produce streak artefacts in the reconstruction. Using 90 views the system delivers a central dose of 90 cGy, and has a density resolution of 4%.Keywords
This publication has 25 references indexed in Scilit:
- Photon mass attenuation and energy-absorption coefficientsPublished by Elsevier ,2002
- Comparison of x‐ray cross sections for diagnostic and therapeutic medical physicsMedical Physics, 1996
- The dose response relationship of a liquid-filled electronic portal imaging deviceMedical Physics, 1996
- Cone-beam CT for radiotherapy applicationsPhysics in Medicine & Biology, 1995
- An analytical edge spread function model for computer fitting and subsequent calculation of the LSF and MTFMedical Physics, 1994
- A review of electronic portal imaging devices (EPIDs)Medical Physics, 1992
- Radiotherapeutic computed tomography with scanned photon beamsInternational Journal of Radiation Oncology*Biology*Physics, 1987
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984
- The effects of a finite number of projection angles and finite lateral sampling of projections on the propagation of statistical errors in transverse section reconstructionPhysics in Medicine & Biology, 1977
- Statistical limitations in transaxial tomographyComputers in Biology and Medicine, 1976