Non-invasive assessment of cardiac output during exercise in chronic obstructive pulmonary disease: comparison of the -rebreathing method and electrical impedance cardiography
- 1 November 1997
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 18 (4) , 327-338
- https://doi.org/10.1088/0967-3334/18/4/006
Abstract
In exercise testing of patients with chronic obstructive pulmonary disease (COPD), non-invasive assessment of stroke volume (SV) and cardiac output (CO) would be valuable. Electrical impedance cardiography (EIC) has proved to be a valid and reliable instrument in healthy subjects. In this study it is investigated whether this also applies to patients with COPD. In 19 COPD patients simultaneous SV measurements were performed during steady-state exercise using the -rebreathing method and EIC (using a fixed blood resistivity value ( or EIC-135 and EIC-150) or a haematocrit based (EIC-ht)). Although close correlations were found (overall correlation between -rebreathing and EIC-ht: R = 0.92 for CO, R = 0.79 for SV), SV and CO measured by means of EIC were significantly higher at low-intensity exercise and lower at high-intensity exercise. The mean differences between the -rebreathing method and EIC-ht were 0.55 ml for SV and for CO (overall exercise data). The limits of agreement (2SD of the mean difference) were 24.7 ml for SV and for CO. These figures are comparable to what is found when healthy subjects are studied. CO was closely correlated to oxygen uptake using the -rebreathing as well as the EIC method; the slope of the regression line was closer to what has been reported in the literature with EIC. Results were better with the EIC-ht than with the EIC-135 and EIC-150 methods. It is concluded that EIC is a reliable and valid method for measurements of SV and CO in COPD during exercise. Keywords: cardiac output, exercise, COPD, impedance cardiographyKeywords
This publication has 20 references indexed in Scilit:
- The intra- and interobserver variability of impedance cardiography in patients at rest and during exercisePhysiological Measurement, 1996
- Noninvasive determination of cardiac output in patients with severe airflow limitation.American Journal of Respiratory and Critical Care Medicine, 1996
- Optimalisation of the spot electrode array in impedance cardiographyMedical & Biological Engineering & Computing, 1996
- Comparing Methods of Measurement: An Alternative ApproachAnesthesiology, 1990
- A comparison between cardiac output determined by impedance cardiography and the rebreathing method during exercise in man.The Japanese Journal of Physiology, 1989
- Discharge instabilities initiated by nonuniform laser extraction in electron-beam sustained discharge KrF lasersJournal of Applied Physics, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Cardiac output measured by impedance cardiography during maximal exercise testsCardiovascular Research, 1985
- Measurement of stroke volume with impedance cardiographyClinical Physiology and Functional Imaging, 1982
- The specific resistance of blood at body temperatureMedical & Biological Engineering & Computing, 1973