Abstract
A new method to measure the tip’s true vibrational amplitude in the shear force microscopy is demonstrated. The measurements are based on modifications of the beam diffraction method. In addition to the dither vibration, the equilibrium position of the tip is set to move along the direction of the vibration. The ratio of ac amplitude to the derivative of dc amplitude gives the tip’s vibration. The sensitivity of this method is determined by the size of the laser focal spot times the ratio of the ac to dc signals of the diffracted light.