Automatic, Noninvasive Intradialytic Clearance Measurement
- 1 April 1993
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 16 (4) , 185-191
- https://doi.org/10.1177/039139889301600404
Abstract
The feasibility of non invasive, automatic intradialytic measurement of effective clearance is demonstrated by a mathematical derivation. In a practical setting the conductivity difference between dialysate inlet and outlet of the dialyzer is measured for two different dialysate inlet concentrations. This allows for the calculation of the electrolyte dialysance, which is equal to the effective urea clearance, provided dialysate flow, blood flow and blood electrolyte composition remain unchanged during the measuring intervalKeywords