Contrast optimization of fluid‐attenuated inversion recovery (flair) imaging
- 1 December 1995
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 34 (6) , 868-877
- https://doi.org/10.1002/mrm.1910340612
Abstract
The optimization of contrast is considered for the fluid‐attenuated inversion recovery (FLAIR) MRI pulse sequence, specifically the contrast of multiple sclerosis (MS) to white matter (WM). A performance bound is identified at 1.5 Tesla as that provided using an inversion time (TI) of 2900 ms. It is shown that TR/TI times exceeding 11000/2600 ms provide about 90% of the MS‐WM contrast possible theoretically. The commonly reported TR/TI combination of 6000/2000 provides only about 60%. For TR times exceeding 8000 ms, an echo time (TE) of 140 ms is at or near optimum. Use of TR/TI times less than 9000/2400 lacks efficiency in multisection imaging. Predicted relative contrast performance of TR/TI 11000/2600 versus 6000/2000 was evaluated in seven patients with known MS lesions, and measurements closely matched theoretical predictions. It is strongly recommended that for near optimum contrast and high multisection efficiency, FLAIR should be performed with TR/TI times exceeding 10000/2500 ms.Keywords
This publication has 23 references indexed in Scilit:
- MRI of the brain stem using fluid attenuated inversion recivery pulse sequencesNeuroradiology, 1993
- High Signal Regions in Normal White Matter Shown by Heavily T2-Weighted CSF Nulled IR SequencesJournal of Computer Assisted Tomography, 1992
- Assessment of demyelination, edema, and gliosis by in vivo determination of T1 and T2 in the brain of patients with acute attack of multiple sclerosisMagnetic Resonance in Medicine, 1989
- Precise relaxation time measurements of normal‐appearing white matter in inflammatory central nervous system diseaseMagnetic Resonance in Medicine, 1989
- In vivo determination of T1 and T2 in the brain of patients with severe but stable multiple sclerosisMagnetic Resonance in Medicine, 1988
- Reproducibility of relaxation and spin-density parameters in phantoms and the human brain measured by MR imaging at 1.5TMagnetic Resonance in Medicine, 1986
- A modified saturation‐recovery approximation for multiple spin‐echo pulse sequencesMagnetic Resonance in Medicine, 1986
- Time‐Multiplexed Multislice Inversion‐Recovery Techniques for NMR ImagingMagnetic Resonance in Medicine, 1985
- Mechanisms of Contrast in NMR ImagingJournal of Computer Assisted Tomography, 1984
- Signal, Noise, and Contrast in Nuclear Magnetic Resonance (NMR) ImagingJournal of Computer Assisted Tomography, 1983