Errors in estimating peritoneal fluid by bioelectrical impedance analysis and total body electrical conductivity.
- 1 February 1993
- journal article
- research article
- Published by Taylor & Francis in Journal of the American College of Nutrition
- Vol. 12 (1) , 66-72
- https://doi.org/10.1080/07315724.1993.10718285
Abstract
Whole-body bioelectrical impedance analysis (BIA) and total body electrical conductivity (TOBEC) have been used to estimate body composition and generalized changes in total body water (TBW). The sensitivity of these methods to measure small, rapid, localized changes in body water has not been fully evaluated. We compared the prediction of TBW by whole-body and segmental BIA and TOBEC with deuterium oxide dilution (D2O) in 10 control subjects and 7 renal failure patients receiving continuous ambulatory peritoneal dialysis (CAPD) prior to and after dialysate infusion. Using D2O as the reference method, there was no significant mean residual error between TBW predicted by BIA and TOBEC in controls (−1.2 +/− 1.5 and −0.9 +/− 1.0 kg) and CAPD patients pre-infusion (−1.0 +/− 2.0 and 0.29 +/− 2.01 kg). After infusing 1.9 +/− 0.18 kg dialysate, the mean residual error between change in body weight and the three methods was −0.44 +/− 0.53 kg for D2O (p < 0.1), −1.7 +/− 0.25 kg for BIA (p < 0.0001), and 1.2 +/− 0.4 kg for TOBEC (p < 0.001). Segmental BIA detected a 7.6% reduction in trunkal resistance with no significant change across the limbs, consistent with abdominal fluid accumulation. It is concluded that whole-body BIA underpredicts and TOBEC overpredicts small changes in peritoneal fluids.Keywords
This publication has 17 references indexed in Scilit:
- Is the Impedance index (ht2/R) significant in predicting total body water?The American Journal of Clinical Nutrition, 1992
- Whole-body and segmental bioelectrical-impedance analysis in patients with cirrhosis of the liver: changes after treatment of ascitesThe American Journal of Clinical Nutrition, 1992
- Bioelectric impedance analysis: Experience with male patients with cirrhosisHepatology, 1991
- Field use of D2 18O to measure energy expenditure of soldiers at different energy intakesJournal of Applied Physiology, 1989
- In vivo isotope-fractionation factors and the measurement of deuterium- and oxygen-18-dilution spaces from plasma, urine, saliva, respiratory water vapor, and carbon dioxideThe American Journal of Clinical Nutrition, 1988
- A new TOBEC instrument and procedure for the assessment of body composition: use of Fourier coefficients to predict lean body mass and total body waterThe American Journal of Clinical Nutrition, 1987
- Validation of tetrapolar bioelectrical impedance method to assess human body compositionJournal of Applied Physiology, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Estimation of human body composition by electrical impedance methods: a comparative studyJournal of Applied Physiology, 1985
- CHANGES DURING HEMODIALYSIS IN TOTAL BODY WATER, CARDIAC OUTPUT AND CHEST FLUID AS DETECTED BY BIOELECTRICAL IMPEDANCE ANALYSISAsaio Journal, 1979