Jarvik 2000 Heart
- 13 October 1998
- journal article
- other
- Published by Wolters Kluwer Health in Circulation
- Vol. 98 (15) , 1568-1574
- https://doi.org/10.1161/01.cir.98.15.1568
Abstract
Background —Mechanical bridge to left ventricular recovery is an emerging strategy for the treatment of heart failure. We sought to validate the use of a new intracardiac axial flow impeller pump for this purpose. Methods and Results —The Jarvik 2000 Heart was implanted into 30 sheep to ascertain mechanical reliability, biocompatibility, and hemodynamic function. We attempted but failed to anticoagulate with warfarin. Elective explants with survival were performed in 3 animals to simulate bridge to recovery. Extensive autopsy studies were performed in all other animals. At speeds between 8000 and 12 000 rpm the device pumped up to 8 L/min, captured all mitral flow, and augmented cardiac output with elevation of mean arterial pressure. The pump was silent and hemolysis negligible. Nonpulsatile flow did not adversely affect neurological or renal function. Device removal proved straightforward and safe. A fractured inflow bearing occurred in 1 early model. There were no other pump failures, but power interruption occurred when the sheep chewed the cables or head-butted the percutaneous pedestal. At autopsy, there was no thromboembolism or primary thrombus formation in any device. Pump occlusion occurred in 2 sheep with bacterial endocarditis. One electively explanted pump, previously switched off for 5 months, had no thrombus in the device or vascular graft. Conclusions —The Jarvik 2000 Heart is a major advance in blood-pump technology and increases the scope of mechanical circulatory support. Reliability and ease of removal favor its use for bridge to myocyte recovery, as well as for bridge to transplantation or long-term support.Keywords
This publication has 19 references indexed in Scilit:
- Evolving Costs of Long-Term Left Ventricular Assist Device ImplantationThe Annals of Thoracic Surgery, 1997
- Mechanical Support in Dilated Cardiomyopathy: Signs of Early Left Ventricular RecoveryThe Annals of Thoracic Surgery, 1997
- Implantable Left Ventricular Assist DevicesAnnals of Surgery, 1997
- The Jarvik 2000 oxford system: Increasing the scope of mechanical circulatory supportThe Journal of Thoracic and Cardiovascular Surgery, 1997
- Development of an Implantable Ventricular Assist SystemThe Annals of Thoracic Surgery, 1997
- Improved Left Ventricular Function After Chronic Left Ventricular UnloadingThe Annals of Thoracic Surgery, 1996
- Miniature axial flow pump for ventricular assistance in children and small adultsThe Journal of Thoracic and Cardiovascular Surgery, 1996
- Improved Mortality and Rehabilitation of Transplant Candidates Treated with a Long-Term Implantable Left Ventricular Assist SystemAnnals of Surgery, 1995
- GUIDELINES FOR SPECIALIST TRAINING IN CARDIOLOGYHeart, 1995
- Implantable electrical left ventricular assist systems: Bridge to transplantation and futureThe Annals of Thoracic Surgery, 1989