Image sequence filtering in quantum-limited noise with applications to low-dose fluoroscopy
- 1 January 1993
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 12 (3) , 610-621
- https://doi.org/10.1109/42.241890
Abstract
Clinical angiography requires hundreds of X-ray images, putting the patients and particularly the medical staff at risk. Dosage reduction involves an inevitable sacrifice in image quality. In this work, the latter problem is addressed by first modeling the signal-dependent, Poisson-distributed noise that arises as a result of this dosage reduction. The commonly utilized noise model for single images is shown to be obtainable from the new model. Stochastic temporal filtering techniques are proposed to enhance clinical fluoroscopy sequences corrupted by quantum mottle. The temporal versions of these filters as developed here are more suitable for filtering image sequences, as correlations along the time axis can be utilized. For these dynamic sequences, the problem of displacement field estimation is treated in conjunction with the filtering stage to ensure that the temporal correlations are taken along the direction of motion to prevent object blur.Keywords
This publication has 35 references indexed in Scilit:
- A general framework for frequency domain multi-channel signal processingIEEE Transactions on Image Processing, 1993
- Enhancement of low-dosage cine-angiographic image sequences using a modified expectation maximization algorithmPublished by SPIE-Intl Soc Optical Eng ,1992
- Nonstationary AR modeling and constrained recursive estimation of the displacement fieldIEEE Transactions on Circuits and Systems for Video Technology, 1992
- Spatiotemporal filtering of digital angiographic image sequences corrupted by quantum mottlePublished by SPIE-Intl Soc Optical Eng ,1991
- Least squares restoration of multichannel imagesIEEE Transactions on Signal Processing, 1991
- Translational motion compensation for coronary angiogram sequencesIEEE Transactions on Medical Imaging, 1989
- A pel-recursive Wiener-based displacement estimation algorithmSignal Processing, 1987
- Noise updating repeated Wiener filter and other adaptive noise smoothing filters using local image statisticsApplied Optics, 1986
- Automated Correction of Patient Motion and Gray Values Prior to Subtraction in Digitized AngiographyIEEE Transactions on Medical Imaging, 1984
- Low-dose, high-frame-rate versus regular-dose, low-frame-rate digital subtraction angiography.Radiology, 1983