Consistency in Circle Cell FT-IR Analysis of Aqueous Solutions
- 1 August 1987
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 41 (6) , 1057-1067
- https://doi.org/10.1366/0003702874447950
Abstract
The CIRCLE CELL®, based on the cylindrical internal reflection technique (CIR), has become a popular FT-IR accessory because it allows for easy collection of IR spectra from aqueous solutions. In this paper we describe a detailed study of the quality of quantification possible with the use of aqueous acetone mixtures over a concentration range of 0.01 to 90% (w/w). During a period of four months, the method's sensitivity, accuracy, precision, and time stability were evaluated. With the use of a Nicolet 60SX FT-IR, an MCT-A detector, 500 scans, and a 300-s sampling delay, five sets of triplicate runs, approximately one month apart, were made at each of 10 concentrations. Our results indicate the following conclusions: (1) quantification by peak height gives better precision, whereas peak area gives a more linear response and correlation with Beer's law over a large concentration range; (2) water bands produce spectral artifacts during subtraction, (3) repositioning of the CIRCLE CELL® optical bench between runs and the micro-boat sampling cell between samples critically affects the quantification results; (4) deviations from Beer's law are caused by chemical effects and insufficient resolution errors; and (5) the reliable sensitivity limit is 0.05% w/w. These findings suggest that precise quantitative analysis of aqueous solutions is now a practical method with the use of FT-IR spectroscopy.Keywords
This publication has 9 references indexed in Scilit:
- On-line high-performance liquid chromatography/Fourier tranform infrared spectrometry with normal and reverse phases using an attenuated total reflectance flow cellAnalytical Chemistry, 1985
- Infrared Intensities of Liquids I: Determination of Infrared Optical and Dielectric Constants by FT-IR Using the CIRCLE ATR CellApplied Spectroscopy, 1985
- The Quantitative Analysis of a Model Fermentation BrothApplied Spectroscopy, 1984
- Infrared Spectroscopy of Aqueous Antibiotic SolutionsApplied Spectroscopy, 1984
- Stability of Rapid Scanning InterferometersApplied Spectroscopy, 1982
- Quantitative Aspects of FT-IR Spectroscopy in Industrial ApplicationsPublished by Springer Nature ,1980
- Noise and Finite Register Effects in Infrared Fourier Transform SpectroscopyApplied Spectroscopy, 1976
- Errors in absorbance measurements in infrared Fourier transform spectrometry because of limited instrument resolutionAnalytical Chemistry, 1975
- Accuracy and Optimization of the Two Prism Technique for Calculating the Optical Constants from ATR DataApplied Spectroscopy, 1970