Computer‐aided endoscopic sinus surgery
- 1 July 1998
- journal article
- clinical trial
- Published by Wiley in The Laryngoscope
- Vol. 108 (7) , 949-961
- https://doi.org/10.1097/00005537-199807000-00001
Abstract
Objectives: To examine four different types of computer‐aided endoscopic sinus surgical devices—the ISG Viewing Wand, the ISG infrared OptoTrak, the IGT FlashPoint 5000, and the VTI InstaTrak—with emphasis on their accuracy and ease of use. Study Design: Prospective study utilizing laboratory experiments and intraoperative data collection. Methods: A review of the literature is presented. Patients undergoing endoscopic sinus surgery during the study period were enlisted under FDA protocols with IRB consent. Groups of patients had surgery performed with each of the above devices—except the FlashPoint 5000. Accuracy measurements were recorded, and user and operating staff comments about ease of use were collected. The FlashPoint 5000 was used exclusively in the laboratory setting, where accuracy measurements were obtained on a cadaver skull. Results: The systems all demonstrated accuracy to within 2.00 mm. Ease of use was somewhat variable, but following a learning curve by the surgeon and operating department personnel, all of the units were considered to be relatively user friendly. Conclusions: Computer‐aided endoscopic sinus surgery appears to be the wave of the future. Nevertheless, the modern endoscopic sinus surgeon must have thorough training in the basic anatomy of the paranasal sinuses as well as the various surgical techniques.Keywords
This publication has 37 references indexed in Scilit:
- An optical 3D digitizer for frameless stereotactic surgeryIEEE Computer Graphics and Applications, 1996
- Use of a Frameless, Armless Stereotactic Wand for Brain Tumor Localization with Two-Dimensional and Three-Dimensional NeuroimagingNeurosurgery, 1993
- Three-dimensional digitizer (neuronavigator): New equipment for computed tomography-guided stereotaxic surgerySurgical Neurology, 1987
- A frameless stereotaxic integration of computerized tomographic imaging and the operating microscopeJournal of Neurosurgery, 1986
- An adapter for computed tomography-guided stereotaxisSurgical Neurology, 1985
- Modification of the Gouda frame to allow stereotactic biopsy of the brain using the GE 8800 computed tomographic scannerNeurosurgery, 1983
- Modification of the Gouda Frame to Allow Stereotactic Biopsy of the Brain Using the GE 8800 Computed Tomographic ScannerNeurosurgery, 1983
- Computed tomographic guidance stereotaxis in the management of intracranial mass lesionsNeurosurgery, 1983
- Computer-assisted Stereotactic Laser Microsurgery for the Treatment of Intracranial NeoplasmsNeurosurgery, 1982
- Development of a computerized microstereotaxic method for localization and removal of minute CNS lesions under direct 3-D visionJournal of Neurosurgery, 1980