Dose distributions of X-ray fields as shaped with multileaf collimators
- 1 January 1992
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 37 (1) , 163-173
- https://doi.org/10.1088/0031-9155/37/1/011
Abstract
Multileaf collimators (MLC) with various blade widths were simulated using standard cerrobend blocks, and three-dimensional dose computations were carried out to study the resultant radiation field edges. Film measurements made with 6 and 18 MV x-ray beams were compared with calculations that employed a three-dimensional Fourier convolution. A spatial accuracy of better than 3 mm was found in the 50% isodose line of the penumbral region with a calculation voxel size of 5 mm x 5 mm x 5 mm. The computer simulation was used to study the deviation of the calculated 50% isodose line from the desired geometric field edge using various MLC blade positions. The study suggests that multileaf collimation to the outside of the desired field edge will lead to overdose outside the field, whereas multileaf collimation to the inside of the desired field edge will lead to underdose inside the field. When the direction of travel of the leaves with respect to the field edge is near 45 degrees, the 50% isodose of a multileaf-collimated beam will fall close to the desired edge with no underdose when the leaf corners are allowed to insert into the desired field edge by 1.2 mm for 6 MV x-rays and 1.4 mm for 18 MV x-rays using a 1 cm wide leaf. These blade offsets account for the scattering of photons and electrons in the medium within the penumbral region.Keywords
This publication has 7 references indexed in Scilit:
- On methods of inhomogeneity corrections for photon transportMedical Physics, 1990
- X-ray dose computations in heterogeneous media using 3-dimensional FFT convolutionPhysics in Medicine & Biology, 1990
- Fast Fourier transform convolution calculations of x‐ray isodose distributions in homogeneous mediaMedical Physics, 1989
- Benefits expected from simple conformal radiotherapy in the treatment of pelvic tumoursRadiotherapy and Oncology, 1988
- Broad beam and narrow beam attenuation in Lipowitz's metalMedical Physics, 1987
- Dosimetric Precision Requirements in Radiation TherapyActa Radiologica: Oncology, 1984
- Pair, Triplet, and Total Atomic Cross Sections (and Mass Attenuation Coefficients) for 1 MeV-100 GeV Photons in Elements Z=1 to 100Journal of Physical and Chemical Reference Data, 1980