Flow microfluorometric system for screening gynecologic cytology specimens using propidium iodide-fluorescein isothiocyanate.

Abstract
Seventy cervical cytology specimens have been screened by a xero resolution flow analyzer-sorter using propidium iodide and fluorescein isothiocyanate as fluorochromes for nucleus and cytoplasm, respectively. This system shows a 1% sensitivity for detection of abnormal cells using only crude visual data analysis. Screening of clinical specimens was performed on the instrument with a 5.8% false negative rate and a 11.8% false positive rate by comparison with routine visual cytologic evaluation of the same samples.