A Multi-Counter System for High Precision Carbon-14 Measurements
- 1 January 1980
- journal article
- research article
- Published by Cambridge University Press (CUP) in Radiocarbon
- Vol. 22 (2) , 442-447
- https://doi.org/10.1017/s0033822200009747
Abstract
A new 14C detector system containing nine, independently working, CO2 proportional counters is described. The system is designed for a sufficient measuring capacity at a precision level better than σ = ± 2‰, which requires a counting time of about one week per sample. The size of the installation requires a simple and economic design of counters and electronics. A single anticoincidence shield for all counters consists of five newly developed flat counters. The modern counting rate (52cpm) is sensitively checked by running Heidelberg sodium carbonate standard samples wth a counting rate of about 10 times modern. A microcomputer (DEC PDP-11 /03) is used for data acquisition. Recent developments in laboratory techniques (preparation and gaschromatographic purification of samples) are also reported.Keywords