Instrumental Problems Related to the Measurement of Quantum Efficiency of Sodium Salicylate
- 1 August 1965
- journal article
- Published by Optica Publishing Group in Journal of the Optical Society of America
- Vol. 55 (8) , 944
- https://doi.org/10.1364/josa.55.000944
Abstract
This paper compares various measurements of the absolute quantum efficiency of sodium salicylate and tries to explain the discrepancies between the different determinations that range from 25–99%. The sensitivity of a phototube is commonly assumed to be independent of the angle of incidence. An experiment described here has demonstrated that this is far from true. Since the photocathode angular sensitivity probably varies from tube to tube, great prudence should be shown when using sodium salicylate for absolute intensity determinations. The described photomultiplier-tube effect might also be of significant importance in scintillation counters and other fluorescence studies. The problems discussed here have been carefully avoided by Allison et al. We therefore conclude that their value of the quantum efficiency, which is equal to 99% for layer thicknesses 2–4 mg/cm2, is the most reliable one.Keywords
This publication has 9 references indexed in Scilit:
- Absolute Quantum Yield of Sodium Salicylate*Journal of the Optical Society of America, 1964
- Some Optical Properties of Sodium Salicylate FilmsApplied Optics, 1964
- Measurement of the Intensity of Vacuum-Ultraviolet Light: the Application of Aromatic HydrocarbonsJournal of the Optical Society of America, 1964
- Absolute Fluorescent Quantum Efficiency of Sodium Salicylate*Journal of the Optical Society of America, 1964
- Fatigue Effects in the Luminescent Yield of Sodium SalicylateApplied Optics, 1964
- The variation of the quantum efficiency of sodium salicylate with thickness of materialBritish Journal of Applied Physics, 1964
- A Simple Method for Preparing Layers of Sodium SalicylateApplied Optics, 1963
- Intensity Measurements in the Vacuum UltravioletJournal of the Optical Society of America, 1953
- Fluorescent Sensitized Photomultipliers for Heterochromatic Photometry in the UltravioletJournal of the Optical Society of America, 1951