Calibration of a Slotted Disk Velocity Selector Using Supersonic Molecular Beams
Open Access
- 1 October 1971
- journal article
- Published by Walter de Gruyter GmbH in Zeitschrift für Naturforschung A
- Vol. 26 (10) , 1634-1638
- https://doi.org/10.1515/zna-1971-1011
Abstract
A new calibration method for slotted disk velocity selectors is presented and is compared with other calibration procedures. It is based on the measurement of the velocity distribution of a supersonic molecular beam produced by a room temperature source. The error due to the assumed theoretical distribution and to experimental procedure is discussed. It is shown that the velocity scale is known to within 0.8%.Keywords