A spiral phantom for IMRT and tomotherapy treatment delivery verification
- 9 November 2000
- journal article
- therapeutic interventions
- Published by Wiley in Medical Physics
- Vol. 27 (11) , 2503-2507
- https://doi.org/10.1118/1.1319523
Abstract
A solid water/cylindrical phantom is machined to create a spiral cavity for placing radiographic or radiochromatic film in a spiral configuration. This spiral phantom is used to sample, predict, and measure data in three‐dimension subspace. The predicted data are obtained by projecting the patient plan data on the spiral phantom in the treatment planning software. The measured data are obtained by irradiating the spiral phantom (with film in the spiral cavity) as per the treatment plan. The predicted and measured data are converted to a two‐dimensional matrix and plotted as a spiralogram. Comparison of these predicted and measured spiralograms provides a quantitative comparison and thus validation of treatment delivered as planned. The spiral phantom is a simple, cost‐effective approach to sample 3D data from complexly shaped, intensity modulated or compensated multiple beams. A software script is being written to automate the entire process of projection, data sampling, and comparison. Design aspects and some examples of dose verification are presented. The usefulness of the spiral phantom for intensity modulated radiation therapy and dynamic field shaping are discussed.Keywords
This publication has 16 references indexed in Scilit:
- Multi-Institutional Clinical Trials: 3-D Conformal Radiotherapy Quality AssurancePublished by S. Karger AG ,2015
- Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapyMedical Physics, 1999
- Evaluation of polymer gels and MRI as a 3-D dosimeter for intensity-modulated radiation therapyMedical Physics, 1999
- An investigation into the applications of polymer gel dosimetry in radiotherapyMedical Physics, 1999
- Extrapolated surface dose measurements with radiochromic filmMedical Physics, 1999
- Commissioning and quality assurance of treatment planning computersPublished by Elsevier ,1993
- Dose-response curves for Fricke-infused agarose gels as obtained by nuclear magnetic resonancePhysics in Medicine & Biology, 1990
- A comprehensive three-dimensional radiation treatment planning systemInternational Journal of Radiation Oncology*Biology*Physics, 1988
- Determination of the accuracy of different computer planning systems for treatment with external photon beamsRadiotherapy and Oncology, 1984
- Performance evaluation of computerized treatment planning systems for radiotherapy: external photon beamsInternational Journal of Radiation Oncology*Biology*Physics, 1980