High-resolution gas chromatography with graphite furnace atomic absorption spectrometry as the detection system
- 1 January 1987
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 2 (8) , 801-803
- https://doi.org/10.1039/ja9870200801
Abstract
A high-resolution gas chromatography-graphite furnace atomic absorption spectrometry (HRGC-GFAAS) system, using on-column injection and signal evaluation with a personal computer (PC), is described. The system affords high specificity and selectivity, combined with a low instrumental detection limit, e.g., 8 pg for tetramethyllead and 16 pg for butyltrimethyllead. The reproducibility and stability of the instrumentation are good, giving a relative standard deviation of 10% at the 100 ng ml–1 level and a correlation coefficient of 0.998 for standard solutions of butyltrimethyllead.Keywords