A Comparison of Stress-Induced Porosity Due to Conventional and a Modified Spinal Fixation Device
- 1 February 1994
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Spinal Disorders
- Vol. 7 (1) , 1-11
- https://doi.org/10.1097/00002517-199407010-00001
Abstract
A hypothesis that device-related osteopenia can be reduced by decreasing the rigidity of a fixation device was tested through a canine study. Polymer washers were interposed between the integral nut and plate of the variable spinal plating (VSP) system to reduce its rigidity. A solid fusion was observed 6 months postoperatively in all of the animals using VSP or modified systems. The stabilized segment using both systems showed similar load-displacement behaviors immediately after surgery and 6 months postoperatively. Although not significant, 5.6 and 1.8% decreases in volumetric density of mineralized bone were found in the stabilized segments due to VSP and modified systems, respectively. The modified system also increased bone growth around screws. The new concept of using polymer washers, to decrease rigidity of the fixation device over time, may reduce device-related osteopenia.Keywords