Continuous Analysis of Ortho-Parahydrogen Mixtures

Abstract
A Gow-Mac thermal conductivity unit has been utilized in the construction of a continuous flow analyzer for ortho-parahydrogen mixtures. No pump or vacuum system is required. The unit possesses adequate sensitivity for operation at room temperature, thus eliminating the need for low temperature baths. Parameters which influence reproducibility have been investigated and simple techniques for their control are suggested.

This publication has 1 reference indexed in Scilit: