Influence of Output Capacitor, Electrode and Pulse Width on Power Consumption in Cardiac Pacing‡
- 1 January 1978
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 1 (1) , 16-24
- https://doi.org/10.1111/j.1540-8159.1978.tb03436.x
Abstract
Three different types of unipolar endocardial electrodes--47 in all--were compared in regard to power consumption at stimulation threshold with six different output capacitors and seven pulse widths. Fifteen were conventional large surface electrodes (area 47 mm2); 18 were conventional small surface electrodes (area 6 mm2), and 14 had a specially designed tip with a large area but small active surface of 8 mm2. Pulse widths ranged from 0.15 to 2.0 msec and output capacitors from 1.0 to 22.0 microFarads. All in all about 2,000 measurements were performed. The average current drain to the pacemaker output stage was measured and power consumption was calculated for each electrode--pulse width--output capacitor combination. In all combinations, the two small surface electrodes consumed approximately the same amount of power and, in both cases, significantly less than the larger one. With regard to power economy at stimulation threshold, the pulse width of choice was about 0.5 msec and, furthermore, power consumption decreased with increasing capacitor size. The optimal combination was a small surface electrode, an output capacitor of 22 microF and a pulse width of 0.5 msec.Keywords
This publication has 13 references indexed in Scilit:
- A low-threshold, non-dislocating endocardial electrodeThe Journal of Thoracic and Cardiovascular Surgery, 1976
- Clinical applications of the threshold tracking pacemakerThe American Journal of Cardiology, 1975
- Pulse duration variation and electrode size as factors in pacemaker longevityThe Journal of Thoracic and Cardiovascular Surgery, 1975
- Comparative studies of “state of the art” and presently used clinical cardiac pacemaker electrodesThe Journal of Thoracic and Cardiovascular Surgery, 1974
- Decreasing electrode size and increasing efficiency of cardiac stimulationJournal of Surgical Research, 1971
- POLARIZATION PHENOMENA RELATING TO PHYSIOLOGICAL ELECTRODESAnnals of the New York Academy of Sciences, 1969
- ALTERATION OF PACEMAKER THRESHOLD BY DRUG AND PHYSIOLOGICAL FACTORSAnnals of the New York Academy of Sciences, 1969
- PHYSIOLOGICAL CHANGES IN THRESHOLDAnnals of the New York Academy of Sciences, 1969
- Endocardial threshold of cardiac response as a function of electrode surface areaJournal of Surgical Research, 1968
- Energy consumption for cardiac stimulation as a function of pulse durationJournal of Surgical Research, 1966