The accuracy of the biox 3700 pulse oximeter in patients receiving vasoactive therapy
- 1 August 1991
- journal article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 17 (8) , 484-486
- https://doi.org/10.1007/bf01690773
Abstract
The accuracy of pulse oximetry for the prediction of oxygen saturation of arterial blood in patients with shock has been hardly studied. This study was undertaken to determine if O2 arterial saturation estimated by the Biox 3700 pulse oximeter (SpO2) with an ear probe could reliably substitute for the measurements of O2 saturation (HbO2) with an IL-282 Co-Oximeter in samples of arterial blood obtained from 24 caucasian patients. All patients were mechanically ventilated and needed vasoactive drugs (dopamine and/or dobutamine). Of 24 patients 13 had shock: cardiogenic (n=6) and septic (n=7). The mean difference between HbO2 and SpO2 was 2.49%±4.24, with a 95% confidence interval of 0.7% to 4.3% (p=0.009). There were clinically important differences between both methods since in 9 of 24 patients (37%) SpO2 values were at least 4% lower or greater than HbO2. This disagreement was also apparent in patients with (6/13, 46%) or without shock (3/11, 27%). To conclude, pulse oximetry is not always a sufficiently reliable method to predict HbO2 in patients with or without shock treated with vasoactive drugs.Keywords
This publication has 15 references indexed in Scilit:
- Pulse Oximeter Failure Thresholds in Hypotension and VasoconstrictionAnesthesiology, 1990
- Reliability of Pulse Oximetry in Titrating Supplemental Oxygen Therapy in Ventilator-Dependent PatientsChest, 1990
- Skin Color and Ear OximetryChest, 1989
- Drug Actions at Mammalian Motor Nerve EndingsAnesthesiology, 1987
- Accuracy of Response of Six Pulse Oximeters to Profound HypoxiaAnesthesiology, 1987
- Oximetry—Know Thy LimitsChest, 1987
- Evaluation of the Ohmeda 3700 Pulse OximeterAnesthesiology, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Noninvasive Detection of Profound Arterial Desaturations Using a Pulse Oximetry DeviceAnesthesiology, 1985
- Clinical evaluation of a new finger oximeterCritical Care Medicine, 1984