Improved Detection of Focal Cerebral Blood Flow Changes Using Three-Dimensional Positron Emission Tomography
- 1 July 1993
- journal article
- research article
- Published by SAGE Publications in Journal of Cerebral Blood Flow & Metabolism
- Vol. 13 (4) , 630-638
- https://doi.org/10.1038/jcbfm.1993.81
Abstract
Removal of the interplane septa and configuration of a typical multislice PET scanner to accept all possible coincidence lines of response leads to a fivefold increase in sensitivity. This can be of value in regional CBF studies using bolus 15O-labeled water injections, allowing the injected dose to be reduced by a factor of 4, while maintaining the same number of noise equivalent counts. Thus, for a given cumulative dose limit, four times as many studies can be performed in a single subject. Data from the three-dimensional Hoffman brain phantom, closely matched to count rates seen in human studies, show that for an identical cumulative dose, the noise in subtraction (stimulus minus baseline) images can be reduced by a factor of 2 using three-dimensional data acquisition, with appropriate fractionation of the dose. This improvement is dependent on axial position due to the sensitivity characteristics of three-dimensional scans; however, there is a significant gain in the signal-to-noise ratio (S/N) in all image planes. Studies performed in a human subject demonstrate how the factor of 2 gain in S/N leads to improved detectability of activation sites in PET subtraction images.Keywords
This publication has 16 references indexed in Scilit:
- Rapid Automated Algorithm for Aligning and Reslicing PET ImagesJournal of Computer Assisted Tomography, 1992
- Evaluation of a 3D reconstruction algorithm for multi-slice PET scannersPhysics in Medicine & Biology, 1992
- Fully three-dimensional reconstruction for a PET camera with retractable septaIEEE Transactions on Medical Imaging, 1991
- 3D PET using a Conventional Multislice Tomograph without SeptaJournal of Computer Assisted Tomography, 1991
- Physiological ROI Definition by Image SubtractionJournal of Cerebral Blood Flow & Metabolism, 1991
- Implementation of three-dimensional image reconstruction for multi-ring positron tomographsPhysics in Medicine & Biology, 1990
- 3-D phantom to simulate cerebral blood flow and metabolic images for PETIEEE Transactions on Nuclear Science, 1990
- A study of the possibility of using multi-slice PET systems for 3D imagingIEEE Transactions on Nuclear Science, 1989
- Enhanced Detection of Focal Brain Responses Using Intersubject Averaging and Change-Distribution Analysis of Subtracted PET ImagesJournal of Cerebral Blood Flow & Metabolism, 1988
- Determination of Object Contour from Projections for Attenuation Correction in Cranial Positron Emission TomographyJournal of Computer Assisted Tomography, 1982