Platelet counting in platelet concentrates with various automated hematology analyzers
- 1 September 2007
- journal article
- Published by Wiley in Transfusion
- Vol. 47 (9) , 1651-1657
- https://doi.org/10.1111/j.1537-2995.2007.01337.x
Abstract
Hematology analyzers use impedance, optical, and/or immunologic techniques for counting platelets (PLTs). PLT counting in whole blood has been validated thoroughly; however, this is not the case for PLT counting in PLT concentrates (PCs), in which red cells (RBCs) are absent. Therefore, this study is focused on PLT counting in PCs to study use of ethylenediaminetetraacetate (EDTA), carryover, and accuracy of the analyzers. In total six hematology analyzers (AcT 8, Beckman Coulter; ADVIA 2,120, Bayer; Cell-Dyn 4,000, Abbott; Onyx, Beckman Coulter; K4,500, Sysmex; and XT 2,000i, Sysmex) were tested for PLT counting. PC samples with various PLT concentrations were made (0-1,700 x 10(9)/L) and measured 10 times. Carryover was determined five times. PC samples (1,000 x 10(9) PLTs/L) in EDTA tubes showed significantly higher PLT counts than samples in "dry" tubes for all analyzers except for the Cell-Dyn 4,000 with the impedance technique. Carryover was not more than 0.3 percent for all analyzers. The K4,500 showed the most accurate results, whereas the Cell-Dyn 4,000 with the impedance technique had low accuracy due to an overestimation of more than 20 percent. Most tested analyzers seemed to be suitable for counting PLTs in PCs. All hematology analyzers should be validated for counting PLTs in absence of RBCs as is the case in PCs, in addition to validation of PLT counting in whole blood.Keywords
This publication has 13 references indexed in Scilit:
- Impact of immunological platelet counting (by the platelet/RBC ratio) on haematological practiceCytometry Part B: Clinical Cytometry, 2005
- Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusionBritish Journal of Haematology, 2005
- The platelet count accuracy of platelet concentrates obtained by using automated analysis is influenced by instrument bias and activated platelet componentsVox Sanguinis, 2004
- Evaluation of the Sysmex Xe‐2100® hematology analyzer in hospital useJournal of Clinical Laboratory Analysis, 2003
- Standardisation of platelet counting accuracy in blood banks by reference to an automated immunoplatelet procedure: comparative evaluation of Cell-Dyn CD4000 impedance and optical platelet countsTransfusion and Apheresis Science, 2001
- Platelet Counting by the RBC/Platelet Ratio MethodAmerican Journal of Clinical Pathology, 2001
- An Interlaboratory Study of a Candidate Reference Method for Platelet CountingAmerican Journal of Clinical Pathology, 2001
- Automated CD61 immunoplatelet analysis of thrombocytopenic samplesBritish Journal of Haematology, 2001
- Effect of EDTA on Platelet Count and Other Platelet Parameters in Blood and Blood Components Collected with CPDA‐1Vox Sanguinis, 1991
- Quality monitoring of haemapheresis platelet concentratesBlood Coagulation & Fibrinolysis, 1991