SERI results from the PEP 1987 Summit Round Robin and a comparison of photovoltaic calibration methods
- 1 March 1989
- report
- Published by Office of Scientific and Technical Information (OSTI)
Abstract
The SERI results for the Photovoltaic Solar Energy Project (PEP) 1987 summit round robin are presented and the proposed International Electrotechnical Commission (IEC) technical committee TC82 global calibration method is evaluated in this report. The global calibration method has a +-4% uncertainty in the short-circuit current (I/sub sc/) including random and nonrandom error sources. This uncertainty can be reduced +-3% if the global normal method is used and the direct component of the total irradiance is measured with a primary absolute cavity radiometer. Nearly half of the uncertainty for various primary and secondary PV calibration methods is nonrandom. If a +-4% uncertainty in I/sub sc/ under standard reporting conditions is acceptable, then the global calibration method with spectral corrections is suitable for cells or modules. If a primary reference device with an uncertainty less than +-3% is required, then the global method is unsuitable. The SERI primary direct normal calibration method has a +-1% uncertainty in I/sub sc/ as verified by comparison with primary AMO cells. If a +-1% primary reference cell is used for secondary calibrations, then a +-2% uncertainty could be achieved for reference cells or modules. 37 refs., 29 figs., 42 tabs.Keywords