Cancellation technique of external noise inside a magnetically shielded room used for biomagnetic measurements
- 1 May 2000
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 71 (5) , 2184-2190
- https://doi.org/10.1063/1.1150603
Abstract
First-order gradiometers inside a magnetically shielded room (MSR) were used to cancel magnetic-field noise. However, the magnetic field inside a MSR is distorted when the amount of external noise is large. This distortion is caused by the low-pass filter property of the MSR. Therefore, the time constants of the frequency-dependent attenuation of the MSR vary spatially and this variation must be taken into account. To investigate noise cancellation, we used a multichannel superconducting quantum interference device consisting of four gradiometers measuring a source signal and two gradiometers as a reference. To compensate for the different magnitudes of the gradiometer wave forms, which differed because of slight differences in their pickup-coil cancel rates, we calculated a fitting parameter. The noise-cancellation method consisted of two processes: reduction of ambient noise caused by the differences in the cancel rate of the gradiometers and a gradient magnetic field inside the MSR, and cancellation of wave-form distortion caused by the spatial variation of the time constants inside the MSR. This cancellation method provides additional attenuation of over 20–30 dB in addition to the balance (>46 dB) of a first-order gradiometer. However, the remaining noise, especially a spike (<2 pT) at the beginning of a large ambient noise step, could not be completely canceled. This noise was caused by the slight difference between the time constants at the reference sensor position and at the signal sensor position. Except for this noise spike, however, the noise cancellation enabled clear magnetocardiogram wave forms to be measured without being affected by strong external noise.Keywords
This publication has 8 references indexed in Scilit:
- A vector fetal magnetocardiogram system with high sensitivityReview of Scientific Instruments, 1999
- Sensitivity of foetal magnetocardiograms versus gestation week.Medical & Biological Engineering & Computing, 1999
- Active shielding to reduce low frequency disturbances in direct current near biomagnetic measurementsReview of Scientific Instruments, 1999
- Evaluation of a high-performance magnetically shielded room for biomagnetic measurementIEEE Transactions on Magnetics, 1996
- Improvement of the performance of a mu -metal magnetically shielded room by means of active compensation (biomagnetic applications)Measurement Science and Technology, 1991
- Spatial discrimination in SQUID gradiometers and 3rd order gradiometer performanceCanadian Journal of Physics, 1982
- Design, construction, and performance of a large-volume magnetic shieldIEEE Transactions on Magnetics, 1982
- Large-volume conventional magnetic shieldsRevue de Physique Appliquée, 1970