Abstract
A comparison between contrast and signal‐to‐noise ratio (SNR), as obtained by both maximum entropy (ME) and 2DFT reconstruction, is reported for NMR images. Our results show that in the ME elaboration the contrast and SNR are generally increased with respect to the 2DFT elaboration. The magnitude of such an increase depends on the ME statistical parameter and also, quite strongly, on the values of contrast and SNR of the 2DFT image. © 1988 Academic Press, Inc.Keywords
This publication has 6 references indexed in Scilit:
- Biological tissue simulation and standard testing material for MRIMagnetic Resonance in Medicine, 1987
- Maximum entropy method in phase‐encoding NMR imagingMagnetic Resonance in Medicine, 1987
- Signal, Noise, and Contrast in Nuclear Magnetic Resonance (NMR) ImagingJournal of Computer Assisted Tomography, 1983
- Image reconstruction from incomplete and noisy dataNature, 1978
- Information Theory and Statistical Mechanics. IIPhysical Review B, 1957
- Information Theory and Statistical MechanicsPhysical Review B, 1957