Reduction in radiation exposure during coronary angiography
- 1 November 1990
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Diagnosis
- Vol. 21 (3) , 195-197
- https://doi.org/10.1002/ccd.1810210314
Abstract
In addition to lead shielding, increased distance between the operator and x‐ray source will lower radiation exposure. To utilize this principle, we interposed a 24 in. piece of pressure tubing between the catheter used for coronary angiography and the manifold apparatus. Radiation exposure to the hand of the operator during coronary angiography was compared with and without the extension tubing. When corrected for the differences in exposure time, operator exposure was 5.38 mrem/min without the extension and 4.84 mrem/min with the extension. Although this is a small difference in exposure/min, a substantial reduction in exposure could accumulate over a 1 yr period. Insertion of this extension tube into the catheter system is a simple and safe way to further reduce operator exposure during coronary angiography.Keywords
This publication has 5 references indexed in Scilit:
- Radiation exposure during diagnostic catheterization and single- and double-vessel percutaneous transluminal coronary angioplastyThe American Journal of Cardiology, 1987
- Radiation exposure and protection in cardiac catheterization laboratoriesThe American Journal of Cardiology, 1985
- Personnel exposure to radiation at some angiographic procedures.Radiology, 1981
- Physician and patient exposure during cardiac catheterization.Circulation, 1978
- Evaluation of Physician Exposure During Cardiac CatheterizationRadiology, 1971