Abstract
Data taken on the temperature dependence of A150 and graphite ionization chambers show a temperature dependence consistent with the volume coefficient of expansion of the chamber materials. For graphite chambers this effect is negligible, but for A150 the T dependence is some 20% less than expected from the ideal gas law. In cases when the highest accuracy is required this effect should be taken into account.Keywords
This publication has 1 reference indexed in Scilit:
- Charge-balancing current integrator with large dynamic rangeNuclear Instruments and Methods in Physics Research, 1983