Response of very small (2 mm) porcine coronary arteries to balloon angioplasty and stent implantation
Open Access
- 13 February 2004
- Vol. 90 (3) , 324-327
- https://doi.org/10.1136/hrt.2003.015305
Abstract
Objective: To investigate the response of very small coronary arteries to stent deployment and balloon angioplasty.Setting: Normal porcine coronary arteries.Methods: 24 pigs underwent intervention to two main coronary arteries, in segments 2.0 mm in diameter, with balloons whose diameter was 2.5 mm at standard pressure. Twelve arteries received a BiodivYsio small vessel (SV) stent; 12 an NIR SV stent; 12 standard BiodivYsio stent, and 12 balloon only. The arteries were harvested at 28 days, fixed, embedded in plastic, and cut and ground in cross section. The injury score and histomorphometry were assessed.Results: The BiodivYsio SV stent was associated with 20% less injury (p = 0.16), a 30% larger lumen (p = 0.13), and a 25% smaller neointima (p = 0.03) than the NIR SV stent, despite identical oversize. The standard BiodivYsio stent exhibited less recoil but 29% greater injury (p = 0.01), 59% more neointima (p = 0.00), and 18% less lumen (p = 0.27) than the BiodivYsio SV. Of all interventions, balloon only was associated with little injury, little neointima, major vessel shrinkage, and the largest lumen.Conclusion: Despite uniform oversize dilatation, both injury and response varied widely in very small porcine coronary arteries, depending on whether a stent or balloon was used, the stent design, and the number and/or thickness of struts. The response to different stent designs is considerable and is related to the degree of injury.Keywords
This publication has 5 references indexed in Scilit:
- Coronary artery stretch versus deep injury in the development of in-stent neointimaHeart, 2002
- Phosphorylcholine-coated stents in porcine coronary arteries: in vivo assessment of biocompatibility.2001
- Intravascular stents: a new technique for tissue processing for histology, immunohistochemistry, and transmission electron microscopyHeart, 1998
- Randomised comparison of implantation of heparin-coated stents with balloon angioplasty in selected patients with coronary artery disease (Benestent II)The Lancet, 1998
- Restenosis and the proportional neointimal response to coronary artery injury: Results in a porcine modelJournal of the American College of Cardiology, 1992