Factors determining the performance of triple quadrupole, quadrupole ion trap and sector field mass spectrometer in electrospray ionization mass spectrometry. 2. Suitability for de novo sequencing
- 31 May 2001
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 15 (13) , 1053-1060
- https://doi.org/10.1002/rcm.338
Abstract
The sequence coverage by fragment ions resulting from collision‐induced dissociation in a triple stage quadrupole (TSQ) and a quadrupole ion trap (QIT) mass spectrometer of 10–20‐mer oligonucleotides was investigated. While (a‐B) and w ion series were the most abundant on both instruments, additional ion series of sequence relevance were preferably formed in the TSQ. Thus, a total number of 83 fragment ions were used to deduce the complete sequence of a 10‐mer oligonucleotide of mixed sequence from a tandem mass spectrum recorded on the TSQ. The complete sequence was also encoded in the 28 fragments that were obtained from the QIT under comparable fragmentation conditions. Spectrum complexity increased considerably at the cost of signal‐to‐noise ratio upon fragmentation of a 20‐mer oligonucleotide in the TSQ, whereas spectrum interpretation with longer oligonucleotides was significantly more straightforward in spectra recorded on the QIT. The extent of fragmentation had to be optimized by appropriate setting of collision energy and choice of precursor ion charge state in order to obtain full sequence coverage by fragments for de novo sequencing. Moreover, full sequence information was also dependent on base sequence because of the low tendency of backbone cleavage at thymidines. Tandem mass spectrometry on the QIT yielded redundant information that was successfully utilized to deduce the complete sequence of 20‐mer oligonucleotides with high confidence. Copyright © 2001 John Wiley & Sons, Ltd.Keywords
This publication has 16 references indexed in Scilit:
- Fragmentation mechanisms of oligodeoxynucleotides studied by H/D exchange and electrospray ionization tandem mass spectrometryJournal of the American Society for Mass Spectrometry, 2001
- Structure and fragmentation mechanisms of isomeric T-rich oligodeoxynucleotides: A comparison of four tandem mass spectrometric methodsJournal of the American Society for Mass Spectrometry, 1998
- Interpretation of Oligonucleotide Mass Spectra for Determination of Sequence Using Electrospray Ionization and Tandem Mass SpectrometryAnalytical Chemistry, 1996
- Sequence Information from 42−108-mer DNAs (Complete for a 50-mer) by Tandem Mass SpectrometryJournal of the American Chemical Society, 1996
- Charged vs. neutral nucleobase loss from multiply charged oligonucleotide anionsJournal of Mass Spectrometry, 1995
- Mass and sequence verification of modified oligonucleotides using electrospray tandem mass spectrometryJournal of Mass Spectrometry, 1995
- Verification of 50- to 100-mer DNA and RNA sequences with high-resolution mass spectrometry.Proceedings of the National Academy of Sciences, 1995
- Ion trap tandem mass spectrometry applied to small multiply charged oligonucleotides with a modified baseJournal of the American Society for Mass Spectrometry, 1994
- Rapid Sequencing of Oligonucleotides by High-Resolution Mass SpectrometryJournal of the American Chemical Society, 1994
- Tandem Mass Spectrometry of Small, Multiply Charged OligonucleotidesJournal of the American Society for Mass Spectrometry, 1992