Comparison of Binax Legionella Urinary Antigen EIA kit with Binax RIA Urinary Antigen kit for detection of Legionella pneumophila serogroup 1 antigen
- 1 June 1996
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 34 (6) , 1579-80
- https://doi.org/10.1128/jcm.34.6.1579-1580.1996
Abstract
The Legionella Urinary Antigen EIA kit (Binax, Portland, Maine) was compared with the EQUATE RIA Legionella Urinary Antigen kit (Binax) for its ability to detect the presence of urinary antigens to Legionella pneumophila serogroup 1. Urine specimens from patients without Legionnaires' disease (n = 33) were negative by both methods (specificity, 100%). Twenty (77%) of 26 urine specimens from patients with Legionnaires' disease positive by the radioimmunoassay kit were also positive by the enzyme immunoassay (EIA) kit. If the cutoff for a positive EIA result were lowered to a ration of > or = 2.5, 23 of 26 (88%) urine specimens would have been positive by EIA and the specificity would remain 100%. Use of the EIA kit is an acceptable method for detecting L. pneumophila serogroup 1 urinary antigens by laboratories that do not want to handle radioactive materials.Keywords
This publication has 5 references indexed in Scilit:
- Reevaluation of the Definition of Legionnaires' Disease: Use of the Urinary Antigen AssayClinical Infectious Diseases, 1995
- Surveillance for Legionnaires' disease. Risk factors for morbidity and mortality.1994
- Reactivity of Serum from Patients with Suspected Legionellosis Against 29 Antigens of Legionellaceae and Legionella-Like Organisms by Indirect Immunofluorescence AssayThe Journal of Infectious Diseases, 1983
- Rapid Radioimmunoassay Diagnosis of Legionnaires' DiseaseAnnals of Internal Medicine, 1981
- Legionnaires' DiseaseNew England Journal of Medicine, 1977