Abstract
A closed feedback loop for the zero velocity drift of a constant acceleration Mössbauer spectrometer has been developed. Using a transducer with a built‐in limit switch and the modified logic circuitry of a multi‐scaler‐analyzer, the feedback loop responds to the zero velocity drift of the entire system as a whole rather than to each component independently. A long‐term zero velocity stability of ±0.001 mm/sec has been obtained.Keywords
This publication has 7 references indexed in Scilit:
- Effect of Zero-Point Motion on the Mössbauer Spectra of K4Fe(CN)6 and K4Fe(CN)6·3H2OThe Journal of Chemical Physics, 1966
- A constant accelerator fly-back Mössbauer spectrometerNuclear Instruments and Methods, 1966
- Mössbauer Effect in Ferroelectric BaTiPhysical Review B, 1965
- Drift-Free Mössbauer SpectrometerReview of Scientific Instruments, 1965
- High Pressure, Low Temperature, Differential-Pressure ProbeReview of Scientific Instruments, 1964
- Inexpensive Automatic Recording Mössbauer SpectrometerReview of Scientific Instruments, 1964
- Mössbauer Effect in Metallic IronPhysical Review B, 1962