A Dynamic Method for Mercury Vapor Detector Calibration
- 1 July 1965
- journal article
- research article
- Published by Taylor & Francis in Aihaj Journal
- Vol. 26 (4) , 388-393
- https://doi.org/10.1080/00028896509342746
Abstract
An apparatus has been designed to facilitate the calibration of mercury vapor detectors. Known volumes of pure air, and air saturated with mercury, are mixed and introduced into a Lucite box which houses the meter under study. Calibration points are then obtained by adjusting the mercury vapor and air dilution ratios. Mercury concentrations are checked chemically by collecting the vapor in acidified potassium permanganate and analyzing by a dithizone extraction method.Keywords
This publication has 3 references indexed in Scilit:
- New Absolute Manometer for Vapor Pressures in the Micron RangeReview of Scientific Instruments, 1955
- Über eine neue Methode zur Messung kleinster Dampfdrucke. Die Dampfdrucke von Quecksilber und KaliumThe European Physical Journal A, 1931
- Measurement of Mercury Vapor Pressure by Means of the Knudsen Pressure GaugePhysical Review B, 1922